|
Europhysics
News (2001) Vol. 32 No. 2
Absorption
and phase imaging with synchrotron radiation
Peter Cloetens1, Elodie Boller1, Wolfgang
Ludwig1, José Baruchel1 and Michel Schlenker2
1 European Synchrotron Radiation Facility, B.P. 220, F-38043
Grenoble, France
2 Laboratoire Louis Néel du CNRS, B.P. 166,
F-38042 Grenoble, France
All users agree on two essential
virtues of synchrotron radiation: its high intensity in the X-ray range,
and its continuous spectrum. The new, third generation, sources have
another, at first sight less spectacular feature: the small divergence
of the beam as seen from the sample. This characteristic is due to the
very small cross-sectional area of the electron beam that acts as the
source of radiation, and to the large source - sample distance. All
three of these qualities lead to novel possibilities, among others in
X-ray imaging. We will discuss some of the approaches developed in hard
X-ray synchrotron radiation imaging, and some of its applications in
quite diverse fields.
Synchrotron radiation refers to the electromagnetic
radiation emitted by ultrarelativistic electrons (energies of several
GeV), circulating in storage rings, at those parts of the rings where
they are accelerated by a magnetic field. This can be uniform over a
part of the trajectory in the bending magnets, or spatially oscillating
in the "insertion devices". The spectrum of the light thus produced
extends from the infra red into the X-ray range, the latter part being
to most users the more valuable one. Emission is strongly concentrated
in the forward direction with respect to the velocity of the emitting
electrons, the characteristic angular opening being mc2/E,
with E the energy of the electrons, m their rest mass. Three
machines in the world belong to the category of third generation, high
energy sources : the European Synchrotron Radiation Facility (ESRF)
in Grenoble, France, at 6 GeV; the Advanced Photon Source in Argonne,
IL, USA at 7 Gev; and SPRING-8 in Japan, at 8 GeV. They are characterised
by the thinness of the electron beam that produces the radiation (source
dimensions < 0.1 mm), and by the provision, in between the bending magnets,
of many straight sections. These allow the positioning of insertion
devices, viz. wigglers or undulators, which can provide each experiment
with the best suited beam.
Imaging is normally associated, in our minds, with
lenses. Unlike visible light or electrons, efficient lenses are not
(yet?) available for hard X-rays, essentially because they interact
weakly with matter, resulting in a refractive index very close (to within
10-5 or 10-6) to unity. Nevertheless X-ray imaging plays an immense
role. Radiographs of hands of attendants made at the lectures on X-rays
in 1896 are the historical banner of X-rays, while each of us benefited
from medical radiography and enjoys the leak-tightness which industrial
radiography controls in pipelines. X-ray tomography is used on a routine
basis, under the name of computed axial tomography (CAT) or medical
scanner, to visualise virtual cuts through human anatomy, obtained from
attenuation measurements performed under different viewing angles. In
radiography (two-dimensional images) as well as in tomography (three-dimensional
exploration), only contrast associated with local variations in X-ray
absorption was, until recently, considered. On the other hand, a different
approach, based on Bragg diffraction ("X-ray topography"), reveals isolated
defects, such as dislocations, inclusions, stacking faults, and sometimes
domain walls, in single crystals. It largely contributed to the development
of the processes that now provide the huge perfect crystals of silicon
which make microelectronics so efficient.
The advent of synchrotron radiation was a boon to
X-ray imaging through the possibilities for real-time observation or
for refined beam preparation offered by the increase in intensity. However,
the place imaging techniques have taken (15% of the activity of ESRF,
i.e. much more than planned) on third generation sources also results
from the entirely new avenues that have been opened through the high
geometrical quality of the beams.
In the following, we separate the developments we
consider as more closely associated to the gain in intensity (the quantitative
aspect) from those more directly connected to the new geometric features
(qualitative aspects). The results we describe were obtained on several
instruments, mostly but not exclusively at ESRF.
Quantitative
progress: absorption imaging ever more refined
Angiography
X-ray absorption changes abruptly with the energy or wavelength
of the photons at the absorption edges, corresponding to the energy
required to eject an electron from a deep atomic level. This variation
has been used for the selective observation of blood vessels, in particular
for the diagnosis of heart problems associated to the coronary arteries,
via angiography. In the hospital version of this technique, a catheter
is introduced along an artery to bring a contrast agent (an iodine-containing
solution) near the coronary artery to be inspected. This approach involves
a fatal risk of about 0.1%. SR makes it possible to obtain excellent
images with reduced danger, because an injection of a contrast agent
through a vein is enough. The higher dilution is compensated by the
logarithmic subtraction of two images. Because the heart moves, the
approach consists in simultaneously using two flat beams with slightly
different energies, one above, the other below the K absorption edge
of iodine (33.17 keV). The beams cross at the location of the investigated
organ, and they are measured by two detector lines, while the patient
is translated vertically. Fig. 1 shows the results recently obtained,
after preliminary attempts on pigs, on human patients at the medical
beamline ID17 of the ESRF.
|
|
Fig 1 Image
of the heart and arteries of the first human patient at ESRF's Medical
Beamline: this differential angiogram results from logarithmic subtraction
of two images obtained simultaneously on either side of the iodine
absorption edge. The right coronary artery (RCA) shows a stent and
the crux (courtesy H. Elleaume, ESRF [H. Elleaume et al., Phys.
Med. Biol. 45, L39-L43 (2000)]). |
Imaging
through magnetic dichroism
Refinement is perhaps even more obvious in the use for imaging
of magnetic X-ray dichroism, which several groups have developed in
the "soft" X-ray range (energy < 1 keV). In a magnetic material, in
the immediate vicinity of an absorption edge, absorption can, for a
given polarisation state of the incident beam, depend on the magnetisation
state of the sample. It is therefore possible to image the distribution
of magnetisation, i.e. the magnetic domains, with only one chemical
element at a time being sensed since the energy of the edge is different
for each. This is obviously attractive in the case of magnetic multilayers,
where the magnetisation directions vary from one layer to the other.
Several approaches are used to change the small variation in X-ray absorption
into an image. Some use an electron-optical system to produce an enlarged
picture through the emitted photoelectrons, which are the more numerous
the higher X-ray absorption is. These experiments, based on the use
of secondary electrons, and on absorption edges corresponding to soft
X-rays, can only be performed in a good vacuum, and only reveal the
immediate neighbourhood of the surface. Another approach is the use
of a transmission X-ray microscope, as at BESSY (Berlin), which proved
a powerful tool for probing the magnetisation process in Gd-Fe multilayers.
|
|
Fig 2 Study
of liquid metal embrittlement by means of high resolution absorption
tomography. Reconstructed slices of a polycrystalline aluminium
alloy: a) in its initial state b) after exposure for 4 hours to
liquid gallium close to room temperature c) after a supplementary
anneal for two hours at 300°C. The gallium appears in b) as
white lines at the grain boundaries. These lines become diffuse
in c) as gallium diffuses into the aluminium grains. The voxel size
in the images is 1 mm, corresponding
to a spatial resolution of 1.7 mm. |
Absorption
microtomography
The three-dimensional (3D) reconstruction of complex structures,
on the basis of many images obtained for different orientations of the
sample with respect to the beam, is possible with spatial resolution
down to 1 mm (microtomography). This is the smaller-scale analogue of
the medical scanner or CAT. From the tomographic reconstruction, one
can at will produce cuts or volume renderings of the object. The number
of 2D images required is approximately equal to the number of pixel
columns in the detector. At the imaging beamline ID19, this number is
typically 900 when using a 1024x1024 pixel CCD camera. The time required
for recording these 2D images with an extended, parallel and moderately
monochromatic (DE/E » 10-2)
beam is now less than 10 minutes, while the 3D reconstruction time,
on a powerful computer, remains on the order of one hour. The spatial
resolution is mainly determined by the detector specially developed
for these experiments. The usual tool is an X-ray --> visible light
converter screen coupled, via visible light optics, to a cooled CCD
camera (FRELON, for Fast REad-out, LOw Noise) developed at ESRF.
|
|
Fig 3 Three-dimensional
renditions of snow samples reconstructed from about 1000 radiographs
obtained at an X-ray energy of 10 keV. The samples were maintained
at a temperature of -60øC in a cryostat during the experiment. The
images correspond to volumes of 3003 voxels with size
10 mm. a) sample of wet snow with well-rounded
grains b) sample with faceted crystals transformed under the action
of a large temperature gradient (of the order of 1°C/cm) c)
melt-freeze crust, partially faceted under a natural temperature
gradient (courtesy C. Coléou). |
Under specific conditions a liquid metal can penetrate
into the grain boundaries of a polycrystalline solid metal, leading
to a brittle behaviour of the normally ductile solid. This phenomenon,
known as liquid metal embrittlement, was discovered more than a century
ago. However, the mechanisms leading to rapid penetration along grain
boundaries remain poorly understood. Absorption radiography and tomography,
with a spatial resolution of about 1 micron, are well suited to studies
on the kinetics of this process in the case of aluminium alloys. Gallium,
which is liquid just above room temperature, attenuates X-rays much
more than aluminium. This makes it possible to observe in-situ the penetration
of gallium into the bulk. Figure 2 shows the same virtual slice of an
aluminium alloy (Al 5038) at various stages. The first tomographic image
(figure 2a) shows the initial state of the sample; the isolated white
points correspond to Fe and Mn rich inclusions. Figure 2b was recorded
after liquid gallium was allowed to penetrate, and after annealing the
sample close to room temperature: the white lines that divide the sample
into cells indicate the presence of gallium along the grain boundaries.
The lines become diffuse and heterogeneous after a second anneal which
allows gallium to diffuse into the grains (figure 2c): isolated gallium
precipitates and cavities can be observed. These experimental observations
should corroborate some of the antagonistic models proposed in the literature.
Figure 3 is a 3D rendition of spring snow that suffered
metamorphic transformations. Metamorphism has radical consequences:
the physical and mechanical properties can change over several orders
of magnitude. This is related to the fact that some snows stick to almost
vertical rock walls, while others yield under the weight of a single
skier. The form and arrangement of the grains, and the quality of the
bonds in the ice, are key factors in determining the properties of snow.
The growth of ice particles is associated with the diffusion of vapour
in dry snow, and to melting / resolidification in damp snow. Dry snows
are usually warmer at the bottom. The value of the temperature gradient
determines whether the crystallites take on a rounded (small gradient)
or facetted (high gradient) form. The important parameters are the area
per unit volume, the inter-grain connections, and the local radius of
curvature of the grains. These parameters cannot be directly measured
on 2D images. 3D images such as those of fig. 3 provide important data
on statistically significant volumes, containing enough grains, and
with spatial resolution limit much smaller than the grain size. They
are quantitatively evaluated by a research group at the Snow Research
Centre (Centre d'Etudes de la Neige) of the French weather agency (Météo-France).
These examples of synchrotron radiation absorption microradiography
show two trends : the use of high spatial resolution in the first case,
and specially designed sample environments in the second. Quantitative
tomography, made possible by the monochromaticity of the X-ray beam,
shifts the emphasis from the sample architecture, i.e. the geometry,
to the densitometry of the solid parts through mapping of the linear
absorption coefficient. Another development is "local" or "zoom" tomography,
the high resolution reconstruction of a region of interest within a
matrix that gets only low resolution reconstruction. This is essential
for applications where it is not possible or not desirable to extract
a small sample from its matrix.
Zone
plate microscopes
The advent of synchrotron radiation also stimulated the development
of microscopes for soft X-rays, in particular for the wavelength range
(the "water window") where absorption by carbon is much stronger than
by water. They are based on make-shifts for lenses, zone-plates, and
reach very valuable performance for biological systems. Thus, 3D reconstruction
of an alga cell was achieved, with a spatial resolution on the order
of 60 nm. The use of zone plates also allows refinements that are usually
thought to be possible only in the visible range, in particular Zernike's
phase contrast technique. A transmission X-ray microscope operating
with energies in the 2- 7 keV range has been installed on the ID21 beamline
of ESRF. The possibility to tune the energy to the vicinity of important
absorption edges (Ca, Cr, ...) allows different elements and their chemical
state to be mapped through spectro-microscopy.
Qualitative
progress: scanning X-ray imaging, phase radiography and phase tomography
Two new families of imaging techniques depend more directly
on the beam quality reached in third generation synchrotron radiation.
They are the scanning imaging techniques on the one hand, and the easy
observation of inhomogeneities in the optical phase of the outgoing
beam on the other hand.
|
|
Fig 4 The
coefficients d and b, representing respectively the phase and absorption,
for aluminium as a function of the X-ray energy. The phase effects
are expected to be dominant in the hard X-ray regimel. |
|
|
Fig 5 Two
images of a hollow sphere consisting of two different polymer layers.
In image a), recorded at a sample-detector distance of 19 cm and
representative of the edge-detection regime, the borders are revealed
by white-black contrast. In image b), recorded at a distance of
3 metres and representative of the holographic regime, many interference
fringes are observed. |
X-ray
scanning imaging
The X-ray beams produced by undulators at third generation
SR sources can, because they originate in a very small source, be focused
into a very fine spot (on the order of a micrometer in linear size)
with extremely high intensity. It is then possible to scan the sample
with respect to the beam, with an appropriate detector or bank of detectors
collecting quantitative information. This information is then displayed
on an electronic device as a point with variable intensity and colour
at the coordinates corresponding to the position of the beam. This provides
the exact analogue of scanning electron microscopy. There is a wide
variety in the type of information available : wide angle diffracted
X-rays, or small-angle scattered X-rays, fluorescence photons, or secondary
electrons. A whole range of scanning X-ray imaging techniques appeared,
and they will certainly develop brilliantly over the next few years.
Why phase?
A beam of X-rays going through an object suffers, apart from
absorption, a phase shift, because the refractive index of materials
is slightly different from unity. The complex refractive index n = 1 - d + i b
describes both absorption (related to b) and the phase shift with respect
to the beam in vacuum:

where the integral is taken along the path of the beam. This phase
is non-uniform over a cross-section of the X- ray beam if the thickness
and/or the structure of the sample are inhomogeneous. d is proportional
to the electronic density of the material, hence approximately to its
mass density; it must be corrected for dispersion when the photon energy
is near an absorption edge. The phase cannot be measured directly, and
variations in phase do not affect the intensity of the beam as it exits
the sample. Phase images have been obtained, using laboratory sources
or second generation synchrotron radiation, using elaborate devices
(interferometers, many- crystal setups). With many images of the sample
recorded while it is rotated in an interferometer, it is possible to
reconstruct in 3D the distribution of refractive index, i.e. to perform
phase tomography.
The qualitatively new character of third generation SR is that its
lateral coherence makes it possible to visualise and reconstruct the
phase variations of hard X-rays, both in simple radiography and in tomography,
with great instrumental simplicity. Furthermore, better spatial
resolution can be reached compared to existing phase imaging techniques.
The required coherence is obtained thanks to the very small angular
size a of the source as seen by a point in the sample (less than a microradian
on the imaging beamline ID19), which entails a large lateral coherence
Lc of the X-ray beam, with wavelength l:

For example, Lc » 100 mm
for l » 1 Å. In optical
terms, the effect used to turn local variations of phase into intensity
variations is interference, at finite distance, between parts of the
beam that have suffered different phase shifts but are coherent with
one another. This is the analogue of defocusing in electron microscopy:
it can be described in terms of Fresnel diffraction, or of in-line holography.
In practice, obtaining a phase-sensitive image just
involves setting the detector (film or CCD camera) at a distance D on
the order of one tenth of a meter from the sample. An absorption image
is obtained if D is small (mm range). This corresponds to the fact that
the region in the sample that affects the image at a point of the detector
has a size r = " (lD) , the radius of the
first Fresnel zone. When D is a few mm, the size of this zone is below
the resolution of the detector (mm), and no interference will be observed:
only absorption contrast will be effective. For larger values of D,
but with r remaining small with respect to the size a of the object
to be imaged, the edges of the object behave independently, and are
the only contribution to the image. The best sensitivity to a phase
object of size a is obtained for a distance D» a2/2l,
but then the image is a hologram which does not look much like the object.
The major advantage of this new type of imaging is the increased sensitivity
it provides. Fig. 4 shows, for aluminium, that d
>> b in the hard X-ray range (E > 6 keV).
This explains the gain observed, particularly for light materials such
as polymers, or for composites consisting of materials which equally
attenuate X-rays, such as Al and SiC.
|
|
Fig 6 a)
Scanning electron microscope image of the surface of a composite
material consisting of an aluminium matrix reinforced by SiC particles
b)-e) Slices obtained by computed tomography corresponding to zones
respectively at the surface, and 25 æm, 50 æm and 75 æm below the
surface. The sample suffered a monotonic tensile test of 1%. Cracked
SiC particles can be detected (arrows A and B), as well as a crack
in the aluminium matrix (arrow C). The tomographic acquisition was
performed at an X-ray energy of 25 keV and a sample- detector distance
of 0.82 m (collaboration with J.-Y. Buffière, INSA, Lyon). |
The phase variations across the beam at the sample
exit lead to variations in intensity, hence to contrast, provided the
phase has a non-vanishing two-dimensional Laplacian, . They show up, for increasing values
of D, through the appearance of a black-white line at the phase jumps
(fig. 5a), then by a set of Fresnel interference fringes which become
more and more obtrusive (fig. 5b): the image is then an in-line hologram.
|
|
Fig 7 Results
of 3D density mapping (holotomography) of human hair. The tomographic
slices in the a) axial and b) radial direction show the medulla
as a low density region within the cortex. Cavities exist along
the medulla and the arrows indicate low density regions. c) 3D rendition
of the hair revealing the medulla within the outer surface. The
3D density mapping involves a phase retrieval method based on images
recorded at several distances (three in this case), a procedure
that is repeated for a large number of angular positions (X-ray
energy 21 keV). |
The set-up used is simple: on the imaging beamline
ID19 at ESRF, it essentially consists in a monochromator made up of
perfect silicon crystals, located about 140 m from the insertion device
(a variable gap wiggler or an undulator) which acts as its photon source.
Of course, phase inhomogeneities give an image also if they do not arise
from the sample. To avoid these unwanted images, the optical elements
in the beamline (beryllium windows, filters and monochromators) are
submitted to unusually stringent demands.
Tomographic
observation through edge-enhancement
The next step consisted in extending this possibility of
observing phase objects to three dimensions, in other words in going
over to phase tomography. It was performed in two steps.
It turns out that the tomographic reconstruction
codes developed for absorption images yield acceptable results when
the phase images result from density jumps with sharp edges. In metallurgy,
X-ray phase microtomography is particularly valuable for systems involving
regions with neighbouring compositions. As shown by fig. 6, it has provided
important results for the understanding of degradation mechanisms in
aluminium-SiC composites. In this case, it is possible both to visualise
the SiC particles easily, and to observe in situ the nucleation
and growth of cracks when the sample is submitted to traction. These
cracks occur first in the elongated particles, and this imaging technique
shows that there are 50% more of these cracks than indicated by investigations
restricted to the surface. The experimental results are now incorporated
in programs that model the mechanical behaviour of these composite materials.
Quantitative
reconstruction of the phase map
The second, recent step consisted in developing a procedure
for the "holographic" reconstruction of the phase maps, which can serve
as the more correct input data for the "tomographic" reconstruction.
Phase retrieval is based on the use of a few images recorded at different
distances from the sample. It therefore leads to acquisition times longer
than for absorption tomography. The approach is derived from a method
originally developed for electron microscopy by a group in Antwerp.
It is quantitative, and determines the phase, well beyond simple edge
images, with a spatial resolution limited by the detector (about 1 mm
in our case). The combination with tomographic reconstruction, called
holotomography, yields 3D quantitative images that show the distribution
of electron density, hence of mass density, in the sample.
Figure 7 shows an example of 3D density mapping obtained
on a human hair. The axial slice (figure 7a) through the middle of the
hair shows that the medulla, the central hair region, is very inhomogeneous
and varies along the hair shaft. The density is in general lower than
the surrounding cortex with some regions revealed as cavities. The radial
slices (figure7b) show low density regions, at the level of the medulla,
but also near the outer border of the hair and in particular points
of the cortex. The lower density of some regions can be ascribed to
a higher lipid concentration and lower protein content.
Conclusion
The advent of third generation synchrotron radiation sources
such as ESRF has opened new possibilities in X-ray imaging. These techniques
are now beyond the stage of demonstration experiments, and they are
appreciated in the solution of various scientific problems in materials
science, biology and medicine. The applications are based, as far as
absorption imaging is concerned, on the very broad choice in photon
energy available (typically between 1 and 120 keV), which makes it possible
to improve the contrast, on the improved spatial resolution (on the
order of a mm), and on the quantitative data evaluation made possible
by the monochromatic and parallel character of the beam. Furthermore,
the very small source size makes scanning imaging techniques possible.
It also provides, in an instrumentally simple way, phase images which
reveal phenomena that are difficult to evidence by other means. This
new technique can be used for two- or three-dimensional imaging, either
qualitatively for edge detection (pores, inclusions...), or quantitatively,
through phase retrieval complemented by tomographic reconstruction,
in the approach called holotomography.
References
[1] Developments in X-Ray Tomography II
Ulrich Bonse, Editor
Proceedings of SPIE Vol. 3772 (1999).
[3] Instrumental aspects of x-ray microbeams in the range above
1 keV
Dhez P., Chevallier P., Lucatorto T.B., Tarrio C.
Rev. Sci. Instrum. 70, 1907-1920 (1999).
[2] Hard X-ray phase imaging using simple propagation of a coherent
synchrotron radiation beam
Cloetens P., Ludwig W., Baruchel J., Guigay J.P., Rejmànkovà-Pernot
P., Salome-Pateyron M., Schlenker M., Buffiere J.Y., Maire E., Peix
G.
J. Phys. D: Appl. Phys. 32, A145-A151 (1999).
[4] Holotomography: Quantitative phase tomography with micrometer
resolution using hard synchrotron radiation X-rays
Cloetens P., Ludwig W., Baruchel J., Van Dyck D., Van Landuyt J. ,
Guigay J.P., Schlenker M.
Appl. Phys. Lett., 75, 2912-2914 (1999).
Copyright EPS
and EDP Sciences,
2001 |